Cleaning
Wash labware as quickly as possible after use. If a thorough cleaning is not
possible immediately, put glassware to soak in water. If labware is not
cleaned immediately, it may become impossible to remove the residue.
Most new glassware is slightly alkaline in
reaction. For precision chemical tests, new glassware should be soaked
several hours in acid water (a 1% solution of hydrochloric or nitric acid)
before washing.
Brushes with wooden or plastic handles are
recommended as they will not scratch or abrade the glass surface.
Glassware
Cleaners
When washing, soap, detergent, or cleaning powder (with or without an
abrasive) may be used. Cleaners for glassware include Alconox®,
Dural®,M&H®, Lux®, Tide® and Fab®. The water should be hot. For glassware
that is exceptionally dirty, a cleaning powder with a mild abrasive action
will give more satisfactory results. The abrasive should not scratch the
glass. During the washing, all parts of the glassware should be thoroughly
scrubbed with a brush. This means that a full set of brushes must be at
hand-brushes to fit large and small test tubes, burets, funnels, graduates
and various sizes of flasks and bottles. Motor driven revolving brushes are
valuable when a large number of tubes or bottles are processed. Do not use
cleaning brushes that are so worn that the spine hits the glass. Serious
scratches may result. Scratched glass is more prone to break during
experiments. Any mark in the uniform surface of glassware is a potential
breaking point, especially when the piece is heated. Do not allow acid to
come into contact with a piece of glassware before the detergent (or soap) is
thoroughly removed. If this happens, a film of grease may be formed.
|
Safe Use of Chromic Acid
If glassware becomes unduly clouded or dirty or contains coagulated organic
matter, it must be cleansed with chromic acid cleaning solution1.
The dichromate should be handled with extreme care because it is a powerful
corrosive and carcinogen.
When chromic acid solution is used the item
may be rinsed with the cleaning solution or it may be filled and allowed to
stand. The length of time it is allowed to stand depends on the amount of
contamination on the glassware. Relatively clean glassware may require only a
few minutes of exposure; if debris is present, such as blood clots, it may be
necessary to let the glassware stand all night. Due to the intense corrosive
action of the chromic acid solution, it is good practice to place the stock
bottle, as well as the glassware being treated, in flat glass pans or pans
made from lead or coated with lead, or plastic polymer pans determined
compatible with the concentration of chromic acid you are using. Extra care
must be taken to be sure chromic acid solution is disposed of properly.
Special types of precipitates may require
removal with nitric acid, aqua regia or fuming sulfuric acid. These are very
corrosive substances and should be used only when required.
|
Rinsing
It is imperative that all soap, detergents and other cleaning fluids be
removed from glassware before use. This is especially important with the detergents,
slight traces of which will interfere with serologic and cultural reactions.
After cleaning, rinse the glassware with
running tap water. When test tubes, graduates, flasks and similar containers
are rinsed with tap water, allow the water to run into and over them for a
short time, then partly fill each piece with water, thoroughly shake and
empty at least six times. Pipets and burets are best rinsed by attaching a
piece of rubber tubing to the faucet and then attaching the delivery end of
the pipets or burets to a hose, allowing the water to run through them. If
the tap water is very hard, it is best to run it through a deionizer before
using.
Rinse the glassware in a large bath of
distilled water. Rinse with distilled water. To conserve distilled water, use
a five gallon bottle as a reservoir. Store it on a shelf near your clean-up
area. Attach a siphon to it and use it for replenishing the reservoir with
used distilled water.
For sensitive microbiologic assays, meticulous
cleaning must be followed by rinsing 12 times in distilled water.
|
Sterilizing Contaminated
Glassware
Glassware which is contaminated with blood clots, such as serology tubes,
culture media, petri dishes, etc., must be sterilized before cleaning. It can
best be processed in the laboratory by placing it in a large bucket or boiler
filled with water, to which 1-2% soft soap or detergent has been added, and
boiled for 30 minutes. The glassware can then be rinsed in tap water,
scrubbed with detergent, rinsed again.
You may autoclave glassware or sterilize it in
large steam ovens or similar apparatus. If viruses or spore-bearing bacteria
are present, autoclaving is absolutely necessary.
|
Handling and Storing
To prevent breakage when rinsing or washing pipets, cylinders or burets, be
careful not to let tips hit the sink or the water tap.
Dry test tubes, culture tubes, flasks and
other labware by hanging them on wooden pegs or placing them in baskets with
their mouths downward and allowing them to dry in the air; or place them in
baskets to dry in an oven2. Drying temperatures should not exceed
140°C. Line the drying basket with a clean cloth to keep the vessel mouths
clean.
Dry burets, pipets and cylinders by standing
them on a folded towel. Protect clean glassware from dust. This is done best
by plugging with cotton, corking, taping a heavy piece of paper over the
mouth or placing the glassware in a dust-free cabinet.
Store glassware in specially designed racks.
Avoid breakage by keeping pieces separated.
Do not store alkaline liquids in volumetric
flasks or burets. Stoppers or stopcocks may stick.
|
Proper care and handling of Pyrex® and PyrexPlus® labware will greatly
increase its life and increase the safety of your work place.
Autoclaving:
PyrexPlus® labware can be
successfully sterilized using liquids or dry cycle sterilization which
involves no vacuum or low vacuum (<5 inches Hg).
Recommended cycles for automated autoclaves
are:
|
Autoclave Type
|
Autoclave
Cycle
|
Gravity
|
Prevacuum
|
Liquid
|
Yes
|
Yes
|
Dry
|
Yes
|
No
|
Prevac
|
-
|
No
|
CAUTION: Always autoclave vessels with loose
caps or closures.
Steam sterilization time should not exceed 15
minutes at 121°C (250°F). Drying time should not exceed 15 minutes at 110°C
(230°F). The actual cavity temperature of the autoclave should be checked to
be sure the autoclave temperature does not exceed the recommended
sterilization and drying temperature.
Autoclaving-Cloudiness:
Should the
coating appear clouded due to dissolved moisture, simply let dry overnight at
room temperature or briefly heat to 110°C (230°F).
Cleaning:
As is common
practice, clean all glassware before use. Any non-abrasive glassware
detergent may be used for hand or automatic dishwasher cleaning. If using a
dishwasher or glassware dryer, care should be taken to be sure the drying
temperature does not exceed 110°C (230°F). Exposure to dry heat should be
minimized.
Avoid brushes and cleaning pads which could
abrade the glass or damage the coating. If using a chromic acid cleaning
solution minimize contact of the solution with the coating.
Labeling and
Marking:
Use
water-based markers for temporary marking or labeling of the PyrexPlus®
labware coating. Solvent-based markers, dyes and stains cannot be removed
from the coating.
NOTE: A slight “plastic” odor may be detected
when handling PyrexPlus® labware. This is due to additives in the plastic
coating which are responsible for its superior performance. The odor is
normal and will not affect the inertness of the inside borosilicate glass
surface.
Burets:
Remove the
stopcock or rubber tip and wash the buret with detergent and water. Rinse
with tap water until all the dirt is removed. Then rinse with distilled water
and dry. Wash the stopcock or rubber tip separately. Before a glass stopcock
is placed in the buret, lubricate the joint with stopcock lubricant. Use only
a small amount of lubricant. Burets should always be covered when not in use.
Culture Tubes:
Culture tubes
which have been used previously must be sterilized before cleaning. The best
method for sterilizing culture tubes is by autoclaving for 30 minutes at
121°C (15 p.s.i. pressure). Media which solidifies on cooling should be
poured out while the tubes are hot. After the tubes are emptied, brush with
detergent and water, rinse thoroughly with tap water, rinse with distilled
water, place in a basket and dry.
If tubes are to be filled with a media which
is sterilized by autoclaving, do not plug until the media is added. Both
media and tubes are thus sterilized with one autoclaving.
If the tubes are to be filled with sterile
media, plug and sterilize the tubes in the autoclave or dry air sterilizer
before adding the media.
Dishes and
Culture Bottles:
Sterilize and
clean as detailed under Culture Tubes. Wrap in heavy paper or place in a
petri dish can. Sterilize in the autoclave or dry air sterilizer.
Pipets:
Place pipets,
tips down, in a cylinder or tall jar of water immediately after use. Do not
drop them into the jar. This may break or chip the tips and render the pipets
useless for accurate measurements. A pad of cotton or glass wool at the
bottom of the jar will help to prevent breaking of the tips. Be certain that
the water level is high enough to immerse the greater portion or all of each
pipet. The pipets may then be drained and placed in a cylinder or jar of
dissolved detergent or, if exceptionally dirty, in a jar of chromic acid
cleaning solution. After soaking for several hours, or overnight, drain the
pipets and run tap water over and through them until all traces of dirt are
removed. Soak the pipets in distilled water for at least one hour. Remove
from the distilled water, rinse, dry the outside with a cloth, shake the
water out and dry.
Blood Cell Count Diluting Pipets:
After use,
rinse thoroughly with cool tap water, distilled water, alcohol, or acetone,
and then ether. Dry by suction. Do not blow into the pipets as this will
cause moisture to condense on the inside of the pipet.
To remove particles of coagulated blood or
dirt, a cleaning solution should be used. One type of solution will suffice
in one case, whereas a stronger solution may be required in another. It is
best to fill the pipet with the cleaning solution and allow to stand
overnight. Sodium hypo chlorite (laundry bleach) or a detergent may be used.
Hydrogen peroxide is also useful. In difficult cases, use concentrated nitric
acid. Some particles may require loosening with a horse hair or piece of fine
wire. Take care not to scratch the inside of the pipet.
Automatic
Pipet Washers:
Where a large
number of pipets are used daily, it is convenient to use an automatic pipet
washer. Some of these, made of metal, can be connected directly by permanent
fixtures to the hot and cold water supplies. Others, such as those made with
polyethylene, can be attached to the water supplies by rubber hose.
Polyethylene baskets and jars may be used for soaking and rinsing pipets in
chromic acid cleaning solution. Electrically heated metallic pipet dryers are
also available.
After drying, place pipets in a dust-free
drawer. Wrap serologic and bacteriologic pipets in paper or place in pipet
cans and sterilize in the dry air sterilizer. Pipets used for transferring
infectious material should have a cotton plug placed in the top end of the
pipet before sterilizing. The plug will prevent the material being measured
from being drawn accidentally into the pipetting device.
Serological
Tubes:
Serological
tubes should be chemically clean, but need not be sterile. However, specimens
of blood which are to be kept for some time at room temperature should be
collected in a sterile container. It may be expedient to sterilize all tubes.
To clean and sterilize tubes containing blood,
discard the clots in a waste container and place the tubes in a large basket.
Put the basket, with others, in a large bucket or boiler. Cover with water,
add a fair quantity of soft soap or detergent and boil for 30 minutes. Rinse
the tubes, clean with a brush, rinse and dry with the usual precautions.
It is imperative when washing serological
glassware that all acids, alkali and detergents be completely removed. Acids,
alkalis and detergents in small amounts interfere with serologic reactions.
Serologic tubes and glassware should be kept separate from all other
glassware and used only for serologic procedures.
Slides and
Cover Glass:
It is
especially important that microscope slides and cover glass used for the
preparation of blood films or bacteriologic smears be perfectly clean and
free from scratches. Slides should be washed, placed in glacial acetic acid
for 10 minutes, rinsed with distilled water and wiped dry with clean paper
towels or cloth. Once the slides have been washed, place them in a wide jar
of alcohol. As needed, remove from the jar and wipe dry. If the slides are
dry stored, wash them with alcohol before use.
|
1. Chromic acid cleaning solution — Use
powdered commercial or technical grade sodium dichromate or potassium
dichromate. If the compound is in the form of crystals, grind to a fine
powder in a mortar. To 20 grams of the powder in a liter beaker, add a little
water, sufficient to make a thin paste. Slowly add approximately 300mL of
commercial concentrated sulfuric acid, stirring well. Transfer to a
glass-stoppered bottle.
Larger amounts can be made in the same
proportions. Use the clear supernatant solution. Chromic acid solution can be
used repeatedly until it begins to turn a greenish color. Dispose of in
accordance with appropriate regulations. Dilute with large volumes of water
before discarding, or carefully neutralize the diluted solution with sodium
hydroxide. Chromic acid solution is strongly acidic and will burn the skin
severely. Use care in handling it.
|